VISCOJET VJ1 SPECIFICATION |
| | |
PERFORMANCE | Range | 0 – 1,000 cP |
| Measurement Time | 1 second |
| Accuracy | 1% of reading or +/- 0.1cP |
| Repeatability | 0.3% of reading or +/- 0.05cP |
OUTPUTS / INPUTS | Analog | 2 × 4-20mA configurable, representing viscosity, ref. viscosity, or temperature |
| Digital | RS485 ModBus Data Link to host system or PC running ViscoLink |
| Power | 24v DC @ 80mA |
TEMPERATURE | Integral Thermometer | Integrated temperature measurement |
| Standard Range | -50 °C to 150 °C |
| Extended Range | To order for greater than 150 °C * |
PRESSURE | Standard Range | 50 Bar |
| Extended Range | To order for greater than 50 bar * |
PIPE CONNECTION | Standard Fitting | 3/4" or 1" BSPP / NPS, 1.5" cone-seat (Swagelok™) |
| Special Fitting | Non-standard process connection to order * |
ELECTRICAL CONNECTION | Cable Length | As required – not affected by cable length up to 1000m |
| Cable Type | Any standard instrument cable |
CONSTRUCTION | All welded, crevice-free |
MATERIALS | Head | SS316 ( EN 1.4401) |
| Sensor | SS316 ( EN 1.4401) |
| Sensor/Fitting Options: | Alloy C22 (EN 2.4602) |
| | Alloy C276 (EN 2.4819) |
| | S32750 Duplex Steel (EN 1.4410) |
| Special Materials: | Non-standard materials to order * |
SURFACE FINISH | Standard | N6 0.8 Ra |
| Special Finish | Non-standard finish options and coatings to order * |
WEIGHT | | 3Kg |
INGRESS PROTECTION | Standard | IP67 |
SOFTWARE | ViscoLink data viewing and logging PC software |
| | |
* SPECIAL OPERATIONS | Custom variations to instrument specification and dimensions routinely provided by our dedicated Special Operations division |
| | |
As it is Hydramotion’s policy to continuously develop and improve its products, some specifications may change without notice |